Aerosol generators of series ATM 220/226 are especially for cleanroom and filter test applications for years highly appreciated by our customers.
Function test at laminar flow boxes (DIN EN 12469), the qualitative test of filters for suspended matter (filter leack test of HEPA and ULPA filters), and acceptance measurements in clean rooms (according to DIN EN ISO 14644 and VDI 2083) are a essential necessity for personal security and product safety.
For such measurements the VDI-directive 3491 proposes an aerosol generator for producing a test aerosol with defined characteristics, like the ATM 220.
By high generated particle numbers at a particle size distribution orientated at the Most Penetration Particle Size (MPPS) of the filter the locating of leaks in the system is possible fast and safe. Using an oily aerosol substance the aerosol generator produces spherical particles mainly in the size range 0.1…0.5 µm with a calculable residence time. That does nearly not extra contaminate the filter material.
For applications where a very precise setting of smaller particle production rates are required (up to 0.6 g / h), the model ATM 221 was developed.
A further range of application for aerosol generators is calibration of measuring devices and visualization of currents. That enables measurements of local velocities and flow profiles of the tracer particles by means of optical methods
Advantages
Applications
Accessories
Principle
Specifications
|
ATM 220 |
ATM 221 |
Compressed air |
max. 8 bar |
max. 2.5 bar |
Volume flow |
max. 250 l/h |
max. 140 l/h |
Mass flow aerosol fluid |
approx. 2.0 g/h |
approx. 0.6 g/h |
Operating time |
max. 25 h |
max. 83h |
Counter pressure |
max. 10 kPa (0,1 bar) |
|
Aerosol outle |
diam. 8 mm |
|
Particle materials |
DEHS, PAO (Emery 3004), Paraffin oil, saline solutions, Latex suspensions,… |
Aerosol Specification for DEHS
Tiêu đề | Link download |
---|---|