Front Keypad, Wall Mountable, Remote Readings
LCD Data Logger gives instant remote readings of temperature, humidity, or atmospheric pressure. Easily started and stopped from a PC. Wall mountable and can store at least 87,000 readings per channel.
Specifications
-
RTD Sensor: Range: -328 to 1562°F (-200 to 850°C); Probe Accuracy: ±0.09°F (±0.05°C) for -328 to 500°F (-200 to 260°C), ±0.5°F (±0.3°C) for 500 to 1562°F (260 to 850°C); Resolution: 0.01°F (0.01°C).
-
Thermocouple Sensor: Internal Range: -4 to 140°F (-20 to 60°C); Accuracy: ±0.9°F (±0.5°C) internally and for types J, K, T, E, and N, ±3°F (±2°C) for types R, S, and B; Resolution: 0.2°F (0.1°C) internally and for types J, K, T, E, and N, 0.9°F (0.5°C) for types R, S, and B.
-
Temperature/Humidity Sensor: Range: -4 to 140°F (-20 to 60°C), 0 to 95% RH; Accuracy: ±0.9°F (±0.5°C) for 32 to 122°F (0 to 50°C), 3% RH; Resolution: 0.2°F (0.1°C), 0.1% RH.
-
Temperature/Humidity/Pressure Sensor: Range: -4 to 140°F (-20 to 60°C), 0 to 95% RH, 0 to 30 psia; Accuracy: ±0.9°F (±0.5°C), ±3% RH, ±1.0% FSR @ 77°F (25°C); Resolution: 0.1°F (0.05°C), 0.1% RH, 0.002 psia.
-
Memory Size: 174,762 readings (DLI-R), 131,071 readings per channel (DLI-T, DLI-H) and 87,381 readings per channel (DLI-P).
-
Sampling Method: Stop on memory full or continuous recording.
-
Sampling Rate: Selectable from 2 sec to 24 hrs.
-
Computer Requirements: Windows® 95, Windows® 98, Windows® 2000, Windows® ME, Windows NT®, and Windows® XP operating system, one free USB port.
-
Power Requirements: 9 V lithium metal battery, installed functional, user replaceable. Optional AC adapter.
-
Battery Life: 1 yr with display off, 30 days with continuous display usage (approx).
-
Interface: USB port (interface cable required).
-
Housing or Enclosure: Black anodized aluminum case.
-
Weight: 16 oz (440 g).
-
Agency Approvals: CE, model DLI-R not CE approved.